• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 6 MM CYLINDRICAL BUR; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE 6 MM CYLINDRICAL BUR; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10036
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, during a cori assisted tka surgery, a real intelligence 6 mm cylindrical bur broke off at the connection.The piece of the broken bur was stuck into the handpiece.No piece fell into the patient.Surgery was finished after a non-significand delay with a sn back-up device.No harm to the patient or further complications reported.
 
Additional Manufacturer Narrative
Section d10 was updated.Section h10: the real intelligence 6 mm cylindrical bur, part number rob10036, lot number 50926756, used for treatment was not returned for evaluation.An image was provided.A relationship between the reported event and the device could not be established.The reported problem could not be confirmed with a visual inspection.The image provided does not confirm the device broke.A complaint history review for similar reported/confirmed complaints concluded this was an isolated event.A review of manufacturing records indicate the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with product mishandling.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE 6 MM CYLINDRICAL BUR
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15664165
Report Number3010266064-2022-00635
Device Sequence Number20429567
Product Code OLO
UDI-Device Identifier00885556757512
UDI-Public00885556757512
Combination Product (Y/N)N
Initial Reporter CountryBE
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2026
Device Model NumberROB10036
Device Catalogue NumberROB10036
Device Lot Number50926756
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date10/24/2022
Supplement Report FDA Received Date11/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/03/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REAL INTELLIGENCE CORI, SERIAL (B)(6).; REAL INTELLIGENCE ROBOTIC DRILL, SERIAL (B)(6).
Patient SexUnknown
-
-