PENUMBRA, INC. POD; HCG, KRD Back to Search Results
Catalog Number RBYPOD3
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure in the left femoral vein using a pod packing coil (pod pc), a pod coil, and a lantern delivery microcatheter (lantern).During the procedure, after advancing approximately half of the pod pc in the lantern, the pod pc became stuck.Therefore, the lantern containing the pod pc was removed.The pod pc was then removed from the lantern and the lantern was reinserted into the patient.Afterwards, while advancing the pod coil through the lantern, the physician broke the pusher assembly of the pod coil.It was reported that the embolization coil was still attached to the pusher assembly.Therefore, the lantern containing the pod coil was removed.The procedure was completed using new pod pcs, and the same lantern.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15664541
Report Number3005168196-2022-00494
Device Sequence Number6498582
Product Code HCG
UDI-Device Identifier00814548019581
UDI-Public814548019581
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/24/2022
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberRBYPOD3
Device Lot NumberF104763
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Initial Report FDA Received Date10/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age23 YR
Patient SexMale
Date Report Sent to FDA01/01/2005
-
-