| Model Number |
SONATA |
| Medical Device Problem Codes |
Impedance Problem (2950); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user experienced a head trauma on (b)(6) 2022, which caused his ear canal to bleed and could not hear sounds with the device.The user has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results revealed overload fractures in the active electrode confirming that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.Other damages found during device investigation are most likely related to the explantation surgery.This is a final report.
|
| |
|
Event or Problem Description
|
|
The user experienced a head trauma on (b)(6) 2022, which caused his ear canal to bleed and could not hear sounds with the device.The user has been re-implanted.
|
| |
|