Model Number 0580-1-441
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/05/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Broken exeter stem.Ceramic head was also revised.
 
Additional Manufacturer Narrative
An event regarding crack/fracture involving a exeter stem was reported.The event was confirmed via evaluation of the returned device.Method & results: -product evaluation and results: visual inspection: visual inspection of the returned device was performed as part of the material analysis: "image shows the exeter v40 stem with fracture location.On visual examination, the fracture was observed through the neck of the stem with the femoral head remaining locked on the stem trunnion." material analysis: a material analysis was performed and concluded the following: "review of exeter v40 stem 44 mm, catalogue # 0580-1-441, lot code g3286940 confirmed a fracture through the neck of the stem.Characterisation using stereo microscopy and scanning electron microscopy confirmed the stem fracture propagated due to fatigue with the final fracture in overload.No manufacturing or material related defects were observed on the device surfaces examined." -clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to fractured stem.A material analysis was performed on the returned device and concluded the following: "review of exeter v40 stem 44 mm, catalogue # 0580-1-441, lot code g3286940 confirmed a fracture through the neck of the stem.Characterisation using stereo microscopy and scanning electron microscopy confirmed the stem fracture propagated due to fatigue with the final fracture in overload.No manufacturing or material related defects were observed on the device surfaces examined." no further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Broken exeter stem.Ceramic head was also revised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXETER V40 STEM 44MM NO 1
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15665070
Report Number0002249697-2022-01544
Device Sequence Number17637159
Product Code LPH
UDI-Device Identifier04546540153302
UDI-Public04546540153302
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2017
Device Model Number0580-1-441
Device Catalogue Number0580-1-441
Device Lot NumberG3286940
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer06/02/2023
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date06/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/19/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-