H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one sealed groshong 8f s/l catheter kit was returned for evaluation.Gross visual evaluation was performed.The investigation is confirmed for the reported tear, rip or hole in device packaging issue, as the top left portion of the product label was noted to be ripped.Evidence of delamination was noted on the outer tray seal.The components were received with a drape inside the kit.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date: 06/2026), g3, h6 (method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|