B. BRAUN AVITUM AG DIALOG®; HIGH PERMEABILIT Back to Search Results
Model Number 710200L
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 05/03/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by the user facility: it was reported that a venous needle dislodged without alarm during treatment.Staff reported the machine never alarmed and blood pump did not stop.The trend file was captured and was provided for evaluation.Upon review of the trend data, it was confirmed that the device did alarm for venous pressure.A safety inspection performed on the device showed no issues.The patient involved in the incident required a blood infusion and was sent to the emergency room for treatment after the event.The customer also confirmed via additional inquiry there was no serious injury and no patient harm was reported after the incident.Estimated time of the occurrence is between one hour and thirteen minutes and one hour and fourteen minutes into the infusion.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical, inc.Internal report number (b)(4).The investigation is ongoing at this time.When additional information does become available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIALOG®
Common Device Name
HIGH PERMEABILIT
Manufacturer (Section D)
B. BRAUN AVITUM AG
schwarzenberger weg 73-79
melsungen, 34212
GM  34212
MDR Report Key15665556
Report Number2521402-2022-00140
Device Sequence Number20625355
Product Code KDI
UDI-Device Identifier04046963686345
UDI-Public(01)04046963686345
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/25/2022
Report Date (Section F) 06/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number710200L
Device Catalogue Number710200L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Device Age9 YR
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/25/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
Date Report Sent to FDA06/01/2022
-
-