MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY HO POR PL HA SZ 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Model Number 71357111
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994)
Date of Event 11/28/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, following a revision surgery on a metal-on-metal tha performed on the patient¿s right hip on (b)(6) 2021 due to severe pain and elevated metal ion levels in blood, the patient presented to the emergency department with right hip pain.A radiographic assessment of the right hip on (b)(6) 2021 revealed acute fractures seen to the right greater trochanter with fracture lines coming to abut the femoral component of the prosthesis.A deformity of the proximal femur was also identified from a remote trauma.Additional details related to the patient¿s clinical course are unknown.
 
Additional Manufacturer Narrative
H3, h6: given the nature of the alleged incident, the device, could not be returned for evaluation.The clinical/medical investigation concluded that, the right hip pain is consistent with the reported acute fractures of the right greater trochanter and deformity of the proximal femur from a remote trauma.However, with the limited information provided the clinical root cause beyond the reported ¿remote trauma¿ cannot be confirmed.It cannot be concluded the reported events were associated with a malperformance of the implant.No subsequent documentation relating to this event was provided, therefore the patient impact beyond the reported events cannot be determined.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the instructions for use documents for total hip systems revealed that congenital deformity, improper implant selection, improper broaching or reaming, osteoporosis, bone defects due to misdirected reaming, trauma, strenuous activity, improper implant alignment or placement, patient non-compliance, can increase risk of femoral or pelvic fractures, also the patient should be advised to report any pain and unusual incidences.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include traumatic injury, postoperative care, patient bone quality and or post-operative patient condition.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY HO POR PL HA SZ 11
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15665759
Report Number1020279-2022-04556
Device Sequence Number2724047
Product Code MRA
UDI-Device Identifier03596010559661
UDI-Public03596010559661
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/13/2019
Device Model Number71357111
Device Catalogue Number71357111
Device Lot Number09AM13681A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/21/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/13/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
R3 3 HOLE ACET SHELL 58MM, LOT#:08MM13322
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-