| Model Number |
20226 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Hematoma (1884)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: an express ld was returned for analysis.A visual examination of the returned device confirmed that the balloon was tightly wrapped and had not been subjected to positive pressure.No issues were noted with the balloon material.A visual examination found the stent was not returned with the device.No issues were noted with the tip of the device.A visual and tactile examination identified no issues along the length of the device.This concludes the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent dislodgement and hematoma occurred.Vascular access was obtained via the femoral artery.The 50% stenosed target lesion was located in the moderately tortuous and mildy calcified right common iliac artery.A 7.0x40x75 cm express ld iliac / biliary stent was advanced for treatment.However, the stent came off the balloon when it was advanced outside the sheath.The device was snared out and another stent was placed successfully.The sheath had to be pulled out and was unable to use a closure device.The patient was noted to develop a hematoma in the groin and had to be transferred to an emergency room.No further intervention was required, and no further patient complications were reported.The patient's status was good.
|
| |
|