• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH REMOVABLE PRN; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH REMOVABLE PRN; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 383335
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported a protruding bulge was found on the bd saf-t-intima¿ safety system with removable prn before use.The following information was provided by the initial reporter, translated from chinese: "when the icu department was ready to use, it was found that there was a protruding bulge on the catheter.".
 
Additional Manufacturer Narrative
Initial reporter phone#: unknown.Initial reporter e-mail: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary since no photos or samples displaying the reported condition of catheter defective / damaged were available for examination, we were unable to fully investigate this incident.Dhr was reviewed and no qns or other events were found related to the complaint stated by the customer.According to sampling plan applied for product performance, this lot was accepted.
 
Event or Problem Description
It was reported a protruding bulge was found on the bd saf-t-intima¿ safety system with removable prn before use.The following information was provided by the initial reporter, translated from chinese: "when the icu department was ready to use, it was found that there was a protruding bulge on the catheter".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SAF-T-INTIMA¿ SAFETY SYSTEM WITH REMOVABLE PRN
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15667302
Report Number9610847-2022-00405
Device Sequence Number12557387
Product Code FOZ
UDI-Device Identifier00382903833351
UDI-Public(01)00382903833351
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013800
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2025
Device Catalogue Number383335
Device Lot Number1140455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer11/03/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-