• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG COAGULATING AND DISSECTING ELECTRODE; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, MONOPOLAR, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG COAGULATING AND DISSECTING ELECTRODE; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, MONOPOLAR, REUSABLE Back to Search Results
Model Number 26775UE
Medical Device Problem Code Melted (1385)
Health Effect - Clinical Code Gastroesophageal Burn (4475)
Date of Event 10/17/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.
 
Event or Problem Description
The event occurred in singapore: it has been reported that the overlap spade and the insulation of the coagulation and dissection electrode melted during activation and caused a burn in the patient.Internal reference number: (b)(4).
 
Event or Problem Description
Internal karl storz reference number: (b)(4).
 
Additional Manufacturer Narrative
According to the customer's statement and the damage pattern, it is likely that the item was used for too long or the hf switch was activated for too long.As a result, the thermal load at the distal end was too high, which led to a burn on the patient and to a melting of the insulation.In addition, the insulation in the second article was found to have a breakdown.This occurs when the peak voltage is above 3.0 kvp.The most probable root cause is a combination of two factors.The first factor is that the hf switch was used for too long, causing heat to build up at the distal end, leading to thermal damage.The second factor was most likely too high a peak voltage, which also contributed to heat generation and insulation breakdown.Therefore, it is concluded that the data/information specified by the ifu was not observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGULATING AND DISSECTING ELECTRODE
Common Device Name
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, MONOPOLAR, REUSABLE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
MDR Report Key15667737
Report Number2020550-2022-00294
Device Sequence Number7179292
Product Code GEI
UDI-Device Identifier04048551072394
UDI-Public4048551072394
Combination Product (Y/N)N
Initial Reporter CountrySN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/15/2022
Report Date (Section F) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26775UE
Device Catalogue Number26775UE
Device Lot NumberOQ03
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/17/2022
Event Location Hospital
Date Report to Manufacturer11/11/2022
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/15/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-