• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 2B 132.5DEG 70MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 2B 132.5DEG 70MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWF602B
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Limb Fracture (4518)
Date of Event 07/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided on a supplemental report.The device remains implanted in the patient.
 
Event or Problem Description
It was reported that the patient had a periprosthetic fracture of the reversed shoulder prosthesis which required reoperation with a osteosynthesis plate.
 
Additional Manufacturer Narrative
The reported event could be confirmed.The device was not returned but evidences were provided, based on provided x-rays which matches the alleged failure (postoperative periprosthetic fracture).Since x-rays and additional data were provided, the opinion of a medical expert was sought and stated as following: "the vancouver b1 periprosthetic fracture is well visible on the x-ray"."yes, it is very likely that the earlier fracture and fracture surgeries may have weakened the bone and made the patient more vulnerable for this fracture.This fracture type is common in traumatic fractures around a well-fixed prosthesis.I have the impression that a former screw hole of an earlier plate may have weakened the bone, leading to this fracture type due to the patient¿s fall".Based on investigation, the root cause was attributed to a patient related issue.The failure was caused by a traumatic event (patient's fall, failure helped by the patient¿s previous medical events).A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that the patient had a periprosthetic fracture of the reversed shoulder prosthesis which required reoperation with a osteosynthesis plate.Update: the fracture occurred following a fall on (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 2B 132.5DEG 70MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR   38330
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15668109
Report Number3000931034-2022-00352
Device Sequence Number12320753
Product Code KWS
UDI-Device Identifier03700434020748
UDI-Public03700434020748
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K122698
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberDWF602B
Device Catalogue NumberDWF602B
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer12/05/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-