MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 100/597/080CZ
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Physical condition of the device showed samples was received decontaminated and without its original packaging.Customer?s reported issue was duplicated.Under visual inspection it was noticed that inflation lumen on the tracheostomy tube was punctured in place where inflation line connects with the lumen.Such defect can happen once adjustable flange is over pulled above the last marking on the tube (close to 15mm connector) which conflicts with instruction for use.The root cause of the reported issue was determined.It is probable that the reported failure occurred during tracheostomy procedure by contact with sharp edge which conflicts the instruction for use.Actions taken to mitigate the reported issue: no corrective actions were taken.During manufacturing process, the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.
 
Event or Problem Description
It was reported that the cuff may have an air leak.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY TUBE
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ  753 01
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
olomoucka 306
hranice 1, mesto 753 0 1
EZ   753 01
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15668399
Report Number3012307300-2022-26163
Device Sequence Number15045968
Product Code BTO
UDI-Device Identifier15019315107146
UDI-Public15019315107146
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K083031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/597/080CZ
Device Catalogue Number100/597/080CZ
Device Lot Number3508304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Initial Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-