• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 4.5MM BROAD LCP® PLATE 14 HOLES/260MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 4.5MM BROAD LCP® PLATE 14 HOLES/260MM; PLATE, FIXATION, BONE Back to Search Results
Model Number 226.641
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported a plate was implanted on an (b)(6) 2022, but broke post-operatively on an unknown date.A revision surgery occurred on (b)(6) 2022 and a new plate implanted.The second plate also broke post-operatively.A second revision surgery occurred on (b)(6) 2022 and the second broken plate was removed.There was no surgical delay and no fragments were left behind in the patient.This complaint captures the rupture of the second plate implanted.Concomitant device reported: unk - screws: locking (part # unknown, lot # unknown, quantity: unknown).This is report 1 of 1 for (b)(4).This report is for an lcp plate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Date of event: unknown when the device broke.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D10: date of concomitant therapy is(b)(6) 2022.H4, h6: part: 226.641, lot: l880094, manufacturing site: werk grenchen, supplier: na, release to warehouse date: 02 may 2018, expiration date: na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo revealed that the lcp 4.5/5 broad 14ho l260 sst had no signs of issue.Due to poor quality photo the complaint condition cannot be confirmed, the device is not shown clearly.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the lcp 4.5/5 broad 14ho l260 sst.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM BROAD LCP® PLATE 14 HOLES/260MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen CO 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15668669
Report Number8030965-2022-08892
Device Sequence Number15046099
Product Code HRS
UDI-Device Identifier10886982159212
UDI-Public(01)10886982159212
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K082807
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number226.641
Device Catalogue Number226.641
Device Lot NumberL880094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer11/07/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexFemale
Patient Weight76 KG
-
-