| Catalog Number |
03P88-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 21-oct-2022, abbott point of care was contacted by a customer regarding i-stat cg8+ cartridges that yielded a suspected discrepant glucose results on a patient.There was no patient information at the time of this report.(b)(6).At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 17-nov-2022.A review of the device history record confirmed the lot passed finished goods release criteria.Retained cartridge testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.
|
| |
|