| Model Number |
C501 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer made sample probe adjustments, performed a reaction cell check, performed a circuit bleach cleaning, and replaced the teflon.Mechanism checks were performed.The customer performed qc successfully.The investigation determined that the cause of the event was consistent with a contaminated and misadjusted sample probe along with insufficient cell wash.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable albt2 tina-quant albumin gen.2 results for 2 patient samples on a cobas 6000 c501 module.For sample 1, the initial albumin result was 77.4 mg/l.The repeat result was 19.3 mg/l.For sample 2, the initial albumin result was 88 mg/l.The repeat result was 41.4 mg/l.No questionable results were reported outside of the laboratory.The reagent lot is 65268801 and the expiration date is 30-apr-2024.
|
| |
|