WILSON-COOK MEDICAL INC 6 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR Back to Search Results
Model Number G25221
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The subassembly history record for the loading catheter said to be involved was reviewed.The quality control tensile test performed on a sample of the manufactured units met the acceptance criteria.A discrepancy or anomaly was not observed with the product that was released for distribution.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.If the knot and the hook are placed closer than 2 cm, this can create resistance when withdrawing the loading catheter through the ligator handle and/or accessory channel of the endoscope.The instructions for use state: ¿attach trigger cord to hook on end of loading catheter, leaving approximately 2 cm of trigger cord between knot and hook.Withdraw loading catheter and trigger cord up through endoscope and out through handle." if additional pressure is applied to the loading catheter when resistance is encountered, this could contribute to separation of the blue hook from the loading catheter.The reported observation can occur if the blue hook gets caught in the accessory channel of the endoscope.The instructions for use advise the user: "it is vital that the integrity of the working channel is intact." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.All loading catheters that are included in the multi-band ligators are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a colonoscopy for hemorrhoidal ligation, the physician used a cook 6 shooter saeed multi-band ligator.It was reported that during the set-up when they were pulling the loading catheter with the trigger cord through the scope, it got stuck around the biopsy cap/handle, they pulled it more and that¿s when the blue hook broke.They try to use the other hook on other end but same thing happened.They end up using hemostat forceps to grab the trigger cord and was able to complete the set up and do the case.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
During a colonoscopy for hemorrhoidal ligation, the physician used a cook 6 shooter saeed multi-band ligator.It was reported that during the set-up when they were pulling the loading catheter with the trigger cord through the scope, it got stuck around the biopsy cap/handle, they pulled it more and that¿s when the blue hook broke.They try to use the other hook on other end but same thing happened.They end up using hemostat forceps to grab the trigger cord and was able to complete the set up and do the case.Three of the bands misfired because of this issue.An unintended section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
See b5 for a correction to the description of event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
MND, LIGATOR
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key15668884
Report Number1037905-2022-00598
Device Sequence Number17744176
Product Code MND
UDI-Device Identifier00827002252219
UDI-Public(01)00827002252219(17)230803(10)W4623642
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2023
Device Model NumberG25221
Device Catalogue NumberMBL-6-XL-C
Device Lot NumberW4623642
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SCOPE UNKNOWN MAKE AND MODEL
Patient Age37 YR
Patient SexFemale
-
-