ARROW INTERNATIONAL LLC CATH PKGD: WEDGE 6 FR 110 CM; CATHETER, FLOW DIRECTED Back to Search Results
Model Number IPN923527
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the balloon did not inflate enough during use.Therefore, it was removed and replaced with another kit of the same product, inserted at the same site to complete the procedure without problem.The user flushed the first sample but did not perform the inflation test before use".The patient health condition is reported as "fine".
 
Additional Manufacturer Narrative
Qn# (b)(4).Other remarks: n/a.Corrected data: n/a.
 
Additional Manufacturer Narrative
Qn#(b)(4).The reported complaint that the "balloon did not inflate enough during use" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.The reported complaint will be monitored for any developing trends.Additionally, according to the event details, the user "did not perform the inflation test before use".As a result, an in-service has been requested to review the instructions for use (ifu) with the customer.Other remarks: n/a.Corrected data: n/a.
 
Event or Problem Description
It was reported that "the balloon did not inflate enough during use.Therefore, it was removed and replaced with another kit of the same product, inserted at the same site to complete the procedure without problem.The user flushed the first sample but did not perform the inflation test before use".The patient health condition is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH PKGD: WEDGE 6 FR 110 CM
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15669052
Report Number3010532612-2022-00456
Device Sequence Number6497513
Product Code DYG
UDI-Device Identifier10801902206746
UDI-Public10801902206746
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K892530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberIPN923527
Device Catalogue NumberAI-07126
Device Lot Number16F22E0093
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-