| Model Number |
1012451-08 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a mildly calcified, mildly tortuous mid left anterior descending artery that is 75% stenosed.A 3.5x12mm xience skypoint was deployed and a 3.5x8mm nc trek was attempted to post-dilate; however, a pinhole rupture was noted during the first inflation at 10 atmospheres.A non-abbott balloon was used to successfully complete the procedure.No additional information was provided.
|
| |
|