• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIMA; TUBE, DRAINAGE, SUPRAPUBIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIMA; TUBE, DRAINAGE, SUPRAPUBIC Back to Search Results
Model Number 32375
Medical Device Problem Codes Contamination (1120); Unsealed Device Packaging (1444)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the package seal and the sterility of the device were compromised.A flexima drainage catheter was selected for use.During preparation, the catheter package was ripped open upon taking out of the box.It was noted that the package seal and the sterility of the device were compromised.The procedure was completed with a different catheter.No patient complications were reported.
 
Event or Problem Description
It was reported that the package seal and the sterility of the device were compromised.A flexima drainage catheter was selected for use.During preparation, the catheter package was ripped open upon taking out of the box.It was noted that the package seal and the sterility of the device were compromised.The procedure was completed with a different catheter.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The costumer attached a picture of the device, and media inspection was performed.It is possible to observe that the pouch was broken so part of the device was exposed through the pouch.The device and the pouch were returned for the analysis, and visual inspection was performed.It was possible to see that the device returned outside the pouch.Also, the pouch was broken, and it had evidence of wrinkles, so it shows that the pouch was manipulated.The catheter, the flexible cannula and a sheath were returned as part of the device, and no damages were observed.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15670518
Report Number2124215-2022-42164
Device Sequence Number20438400
Product Code GBO
UDI-Device Identifier08714729323310
UDI-Public08714729323310
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date05/31/2025
Device Model Number32375
Device Catalogue Number32375
Device Lot Number0029513998
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date12/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-