• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 1013468-020
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat the tibio-peroneal trunk.The 5.0x20mm armada 18 balloon dilatation catheter (bdc) would not deflate with first inflation at 8 atmospheres.Therefore, it was withdrawn while inflated and noted to go through sheath without issue.There were no patient effects and no significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the contrast mix used in the procedure was 70% saline/ 30% contrast.It should be noted that the percutaneous transluminal angioplasty (pta) catheter, armada 18, ce instructions for use specified that the contrast diluted 1:1 with normal saline.In this case, since the contrast ratio used was less than the required percentage it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.2017: contrast incorrect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15670607
Report Number2024168-2022-11040
Device Sequence Number12377477
Product Code LIT
UDI-Device Identifier08717648208164
UDI-Public08717648208164
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model Number1013468-020
Device Catalogue Number1013468-020
Device Lot Number1101941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2022
Initial Report FDA Received Date10/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-