Model Number 434892
Health Effect - Clinical Codes Pericardial Effusion (3271); Thrombosis/Thrombus (4440)
Date of Event 09/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
A coronary angiography was performed including using a shockwave balloon in a svg.The angiogram showed a perforation in the venous graft at the site of the shockwave therapy.An attempt was made to seal the perforation with a pk papyrus covered stent system, but it could not be delivered and the stent was lost (reported separately).The complaint pk papyrus device was deployed but could not seal the 12mm long perforation completely.The patient condition deteriorated, a thrombus was detected as procedure related complication and surgery was discussed but in view of patients history it was decided against it.Patient was shifted to ccu.Device code is (b)(4).
 
Additional Manufacturer Narrative
The affected pk papyrus stent system was not returned to biotronik and could therefore not be subjected to a technical investigation.Images of the case such as angiographies could also not be obtained.The provided clinical information (pk papyrus device registration form and cathlab report) was sufficient to establish whether a device deficiency contributed to this event.The review of the product release documentation confirmed that the device was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.Based on the provided documentation and the conducted review no manufacturing related root cause could be identified.The treating physician rated the outcome not as a device related but as a procedure related complication.Pk papyrus is indicated for the treatment of acute perforations of native coronary arteries and coronary bypass grafts in vessels 2.5 to 5.0 mm in diameter.As such, this device is typically used in emergent response to a life-threatening adverse procedural event unrelated to use of the pk papyrus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PK PAPYRUS US 5.0/15
Common Device Name
COVERED CORONARY STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15670627
Report Number1028232-2022-05460
Device Sequence Number20448760
Product Code NIV
UDI-Device Identifier07640130447202
UDI-Public(01)07640130447202(17)2303
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
H170004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number434892
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03211113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/28/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization; Life Threatening;
Patient SexMale
-
-