| Model Number |
710300 |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was to be sent to a repair facility (acist) for investigation and necessary repairs.Steris is pending the report of the evaluations.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that their co2mpact endoscopic insufflator front panel is not functioning properly resulting in a procedure delay.No patient harm or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of the evaluation determined that the unit's cpu board, the membrane panel, and the lower valve were damaged.The unit was repaired, tested, found to be operating according to specification, and returned to service.No additional issues have been reported.
|
| |
|