| Model Number |
460405 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/17/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
This is 1 of 3 reports and is linked to mfg report numbers: 3014334038-2022-00240, 3014334038-2022-00241.A facility reported having issues with the carbide burs (dfg700l) breaking, immediately.It was reported that this event occurred during dental extraction.Pieces fell into the surgical site and was removed with thumb forceps.Not all broken pieces were recovered; however, no pieces were left inside the patient.No patient injury or significant delay in surgery occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been no return of product after three (3) good faith efforts (gfe) attempts; therefore, an evaluation of the device could not be performed.Lot number information has been provided; device history record (dhr) was reviewed, and no anomalies related to the reported failure were found.A definitive root cause of the reported issue could not be determined.The issue of breaking burs may be the result of misalignment during procedure or environmental damage.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|
Search Alerts/Recalls
|