| Model Number |
453956 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An orsiro mission drug-eluting stent system was selected for treatment of a moderately calcified lesion (80 percent stenosis degree) in the moderately tortuous mid lad.It was attempted to cross a challenging lesion in the lad with orsiro mission stent systems and the stents became dislodged.This happened with two stents of same size and same lot and is more attributable to the properties of the lesion.The stents were removed with no complication.This is two of two.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes neither the complaint instrument nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter and the stent retention force of a defined amount of samples is tested from every lot.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.
|
| |
|