| Model Number |
710300 |
| Medical Device Problem Codes |
Insufficient Device Problem Information (3190); Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was to be sent to a repair facility (acist) for investigation and necessary repairs.Steris is pending the report of the evaluations.A follow-up report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that their co2mpact endoscopic insufflator entire panel is turned on with lights at the half flow resulting in a procedure delay.No patient harm or injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the evaluation performed by the repair facility (acist) determined the unit was out of calibration.The unit was recalibrated, tested, and returned to service.
|
| |
|