Model Number ZA9003
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer indicated that patient information cannot be provided due to personal data privacy legislation/policy.Date of event: the exact date is unknown.The best estimate is between the implant and explant dates: (b)(6) 2016 through (b)(6) 2022.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).Attempts have been made to obtain the missing information.However, to date, it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
It was reported that a non-preloaded intraocular lens (iol) was explanted from the patient¿s left eye due to the lens was displaced post-operatively.Regarding how long the issue persisted, the customer responded it occurred during the time between surgeries.An unplanned vitrectomy was performed.Customer also stated no sutures were required, no incision enlargement, no prescribed medication.They indicated that the directions for use were followed.The replacement iol is the same model lens with the same diopter power.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: nov 03, 2022.Section h3: evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received crushed inside of the lens insert and with a haptic bent.The lens was cleaned revealing, that the lens was damaged from the lens insert.No further issues were identified.A product quality deficiency could not be determined.Manufacturing record review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15671284
Report Number3012236936-2022-02555
Device Sequence Number20438289
Product Code HQL
UDI-Device Identifier05050474528949
UDI-Public(01)05050474528949(17)210304
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2021
Device Model NumberZA9003
Device Catalogue NumberZA90030165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date10/25/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/04/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-