| Catalog Number |
PR20XT |
| Medical Device Problem Codes |
Break (1069); Excessive Heating (4030)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via clinical trial (b)(4) the probe was used for the liver ablation cycle.The placement was normal and no adjacent structures aside from normal liver parenchyma.The cycle shut off prematurely and an error message appeared saying that probe was overheating due to nearby foreign body.The imaging was then reviewed by the clinician and no abnormal findings were reported or immediate threat to the patient discovered.However, when the probe was then removed it was found that the tip (approximately 2cm in length) appeared to have broken off and remained inside the patient embedded in the liver soft tissue (approximately 10cm deep).The patient was then clinically evaluated for safety, but the probe piece was unable to be removed in the intervention radiology setting.The patient will be reviewed and managed by the patient¿s primary care team in the near future.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unknown.Additional information was requested and the following was obtained: was the patient transferred from a intervention radiology suit table to a ct imaging table with the probe in place? the patient was not moved in the time i was there so i am unable to comment on this.If yes, what precautions were used to prevent the probe from moving while transferring the patient? what is the relationship of the adverse event to the study device? i will refer to the physician for this.Query: medical/safety review: please confirm whether there was challenge/resistance encountered or there was any lateral force applied during the probe tip insertion.In the investigator's opinion, what might be the cause or contributing factors for the tip breakage? please enter your response as a query response.Thank you.Site response: no unusual resistance.The probe was placed without any issues and there was nothing noted until after it was turned on.Investigator states that he is unsure of the cause or any contributing factors for the tip breakage.A pr xt 20cm 15 gauge neuwave probe was used for the ablation of a liver lesion.The placement was normal and no adjacent structures aside from normal liver parenchyma.The ablation cycle shut off prematurely and an error message appeared saying that probe was overheating due to a nearby foreign body.The imaging was then reviewed by the ir attending with no abnormal findings.When the probe was removed it was found that the tip had broken off and a piece remained inside the patient, embedded in the liver soft tissue (approximately 10cm deep).Because of it's location in the soft tissue, the detached piece could not be removed in the interventional radiology setting (it would have required a transfer to surgery) however it was determined after thorough discussion and review that the probe did not pose an immediate risk to the patient's health and could remain in place for now.The ablation procedure was completed without incident using a new probe.No direct action was taken at the time beyond standard intraprocedural imaging.An immediate non contrast enhanced ct was taken alongside ct fluoroscopy of the ablation defect to confirm the nature of the product failure as well as location of the probe tip (interventional radiology procedure room that the ablation took place in has a pet scanner and a ct scanner with a ct fluoroscopy functionality, and option for contrast injection, attached to the operating table.All scans were performed inter-procedurally with the patient in the same position) the contrast enhanced ct was reviewed for any evidence of bleeding or damage to adjacent structures, and none were found the patient was not transferred or repositioned during the course of the event.The patient was admitted overnight and discharged the following morning without incident (kept overnight due to the timing of the case and not due to any complications or extended stay).It has been determined that the adverse event is "definitely"/"clearly related" to the device attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing records evaluation could not be performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 12/7/2022.Investigation summary: the call home data revealed reflected power errors and higher than expected ablation temperatures.The probe was returned with a broken probe tip (not included, this tip was left in the patient as stated in the complaint description).The probe was analyzed, and probe displayed evidence of thermal damage.However, the root cause is unknown because of the damage seen, as well as the entire probe not being returned.Photo attached displays a broken probe tip.A search of nonconformities (ncs) was performed for lot ml22047743.There were no observed nonconformities for this lot.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/16/2022 additional information received: see attached mw 5112766.
|
| |
|