| Model Number |
IPN923528 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the user inflated the balloon during use on a patient, but it deflated soon.Therefore, the catheter was removed and replaced with another kit.There was no report of patient complications, serious injury, or death.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|
Event or Problem Description
|
|
It was reported that, the user inflated the balloon during use on a patient, but it deflated soon.Therefore, the catheter was removed and replaced with another kit.There was no report of patient complications, serious injury, or death.
|
| |
|