| Model Number |
N/A |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Concomitant medical products: 20123606- g7 longevity high wall 36mm f-64538079.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2022 - 02427.Customer has indicated that the product will not be returned to zimmer biomet for investigation, product discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported patient underwent a right hip revision approximately 1-day post implantation due to the misuse between the head and liner.Surgeon used a 36mm id liner with a 40mm head instead of a 36mm head after declining to verify implants before being opened onto the field by the circulating nurse.No additional information on the reported event is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; g3; h2; h3; h4; h6 h6 component code: mechanical (g04) - head no product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.A 40mm cocr head was implanted instead of a 36mm thus, the root cause of the reported issue is attributed usage error - misidentified.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional information on the reported event.
|
| |
|