| Model Number |
66-1997 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).510k is unknown.No information has been provided to date.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of product.A product sample was received for evaluation.Visual and functional testing were performed.One sample was received.During the visual inspection, it was observed the sample was missing the patient connector.The root cause was related to the manufacturing process.Due to the low occurrence no action will be taken at this time.Complaints will continue to be monitored for any trends in incorrect assembly.
|
| |
|
Event or Problem Description
|
|
It was reported that during the pre-use check, the customer noticed no patient-side connector came with it the product.No patient injury was reported.
|
| |
|