Model Number 26173KL
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.
 
Event or Problem Description
It was reported that during surgery, a small metal piece is dislocated from the needle holder.This fragment ended up in the patient's abdomen.There is no information the fragment in question could be retrieved.Internal reference number: (b)(4).
 
Additional Manufacturer Narrative
During surgery, a small metal piece is dislocated from the needle holder.This fragment ended up in the patient's abdomen.There is no information the fragment in question could be retrieved.Upon evaluation of the instrument, it could be confirmed that it was bent.In addition, the carbide insert is detached.The most probable root cause is improper handling.Further evidence for this probable root cause is a video provided to us by the hospital.This video shows that the needle was bent straight at the end of the sewing process using two needle holders.After consultation, this is not the correct procedure, but the needle is removed through the trocar in its original state.Bending/rotating the needle holders exerts forces on the inserts, which caused the reported malfunction.In conclusion, the failure could be confirmed, but a manufacturing error can be excluded.The customer will be informed about the evaluation results in a separate letter.Internal karl storz reference number: (b)(4).
 
Event or Problem Description
Internal karl storz reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REUSABLE NEEDLE HOLDER
Common Device Name
REUSABLE NEEDLE HOLDER
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
MDR Report Key15675194
Report Number2020550-2022-00293
Device Sequence Number20426196
Product Code GCJ
UDI-Device Identifier04048551001110
UDI-Public4048551001110
Combination Product (Y/N)N
Initial Reporter CountrySW
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
Report Date (Section F) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number26173KL
Device Catalogue Number26173KL
Device Lot NumberXQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/16/2022
Event Location Hospital
Date Report to Manufacturer11/16/2022
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date11/17/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-