• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE CAGE GLENOID S, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE CAGE GLENOID S, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number EQUINOXE CAGE GLENOID S, POST AUG, RIGHT
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Implant Pain (4561)
Date of Event 09/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitants: 300-10-45 - equinoxe replicator plate 4.5mm o/s.(b)(4).300-20-02 - equinox square torque define screw drive kit.(b)(4).300-30-06 - equinoxe preserve stem 6mm.(b)(4).310-02-44 - equinoxe, humeral head tall, 44mm (alpha) (b)(4).321-20-00 - equinoxe reverse shoulder drill kit.(b)(4).531-78-20 - shoulder gps hex pins kit (b)(4).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported a female patient, initial right shoulder implanted on (b)(6) 2018, underwent a revision procedure on (b)(6) 2022, approximately 3 years 9 months post the initial procedure.The patient complained of pain.The surgeon findings indicated wear on glenoid.Patient last known to be in stable condition.The explants are not available for return.No other information available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: a, b, c, d, e, g.The revision reported was likely the result of prosthesis wear and fracture.An additional reason for the reported revision may be inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID S, POST AUG, RIGHT
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15675281
Report Number1038671-2022-01354
Device Sequence Number17753185
Product Code KWS
UDI-Device Identifier10885862201003
UDI-Public10885862201003
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/30/2021
Device Model NumberEQUINOXE CAGE GLENOID S, POST AUG, RIGHT
Device Catalogue Number314-13-32
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer03/27/2025
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date03/28/2025
Date Device Manufactured02/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-