• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 13.5 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 13.5 CM QD ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC Back to Search Results
Model Number QD14-S-G1
Medical Device Problem Codes Appropriate Device Problem Term/Code Not Available (3191); Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from switzerland that during service and evaluation, it was determined that the attachment device had heat, bearing damage, and the nose tube was bent/damaged.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that during pre-surgery testing it was observed that the device was not working properly.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device was generating heat.Therefore, the reported condition that the device was not working properly was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi ¿ (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
13.5 CM QD ANGLE ATTACHMENT
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key15675368
Report Number1045834-2022-01426
Device Sequence Number20426199
Product Code HBC
UDI-Device Identifier00845384016465
UDI-Public00845384016465
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQD14-S-G1
Device Catalogue NumberQD14-S-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/19/2022
Initial Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/25/2016
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-