Model Number 1013468-200
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.The investigation determined the reported balloon rupture appears to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.The additional armada 18 and armada 14 devices referenced are being filed under separate medwatch report numbers.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the superficial femoral artery (sfa) with moderate calcification and no tortuosity and anterior tibial artery with moderate calcification and no tortuosity.The 5.0x120mm armada 18 balloon dilatation catheter (bdc) advanced without any resistance noted and ruptured at nominal pressure during the first inflation.Another 5.0x200mm armada 18 (bdc) was then advance without resistance noted as well, however, ruptured during the first inflation at 14 atmospheres (atms).The balloon pressure was held for about 2 minutes and at about 1.5 minutes it ruptured.The 2.0x200mm armada 14 bdc wad advanced to the anterior tibial artery without resistance but also ruptured at nominal pressure during the first inflation.Again, the pressure was held for 2 minutes but this time ruptured after about a minute.All devices were prepared per instructions for use (ifu).Non-abbott balloons were used to complete the procedure for the sfa and the anterior tibial artery.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15676054
Report Number2024168-2022-11062
Device Sequence Number20425342
Product Code LIT
UDI-Device Identifier08717648208249
UDI-Public08717648208249
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1013468-200
Device Catalogue Number1013468-200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/21/2022
Initial Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-