| Model Number |
1864G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Product not returned.Additional information was requested, and the following was obtained: attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Please provide the lot number: no further information is available.Device return status: the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.
|
| |
|
Event or Problem Description
|
|
It was reported a patient underwent an unknown surgery on (b)(6) 2022 and a suture was used.The needle tip was broken during use.The needle tip did not fall into the patient.Further details are not provided.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #(b)(4).Date sent to the fda: 12/14/2022.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.H3 investigational summary: the product received for analysis was identified as product code 1864g.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the tip of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 11/18/2022.Additional information was requested, the following was obtained: - please provide the lot number: no further information is available.- device return status: the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|