ETHICON INC. ETHLN BLK 18IN 4-0 S/A PC-3 PRM MP; SUTURE, NONABSORBABLE, SYNTHETIC Back to Search Results
Model Number 1864G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Product not returned.Additional information was requested, and the following was obtained: attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Please provide the lot number: no further information is available.Device return status: the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.
 
Event or Problem Description
It was reported a patient underwent an unknown surgery on (b)(6) 2022 and a suture was used.The needle tip was broken during use.The needle tip did not fall into the patient.Further details are not provided.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Date sent to the fda: 12/14/2022.The lot/batch was not provided; therefore, a manufacturing record evaluation could not be performed.H3 investigational summary: the product received for analysis was identified as product code 1864g.Visual inspection and metallurgical analysis were performed on the returned product.A fracture was observed at the tip of the needle.One side of the needle was received, the mating fracture surface was not provided for this evaluation.The needle was noted with marks that appears to be by surgical instrument.A scanning electron microscope was used to examine the fracture surfaces and surrounding area of the needle.The fracture surfaces were examined in multiple locations to determine the fracture mode.The evaluation of the sample revealed the fracture was composed of microvoid coalescence, which is evidence of a ductile fracture mode.Mechanical damage observed coincidental to the fracture provides additional evidence that the failure was induced by mechanical deformation leading to ductile overload.This was a ductile fracture.The evidence of this examination indicates that the breakage occurred at the tip of the needle during use due to tensile overload.There is no evidence of any material flaw or defect that would cause premature failure.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 11/18/2022.Additional information was requested, the following was obtained: - please provide the lot number: no further information is available.- device return status: the device has been received at sukagawa and will be shipped.Please check rmao.No further information will be provided.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETHLN BLK 18IN 4-0 S/A PC-3 PRM MP
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15676692
Report Number2210968-2022-08800
Device Sequence Number6444011
Product Code GAR
UDI-Device Identifier10705031002814
UDI-Public10705031002814
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1864G
Device Catalogue Number1864G
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer10/31/2022
11/22/2022
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date11/18/2022
12/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-