AMO PUERTO RICO MFG. INC. TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Impairment (2138); Halo (2227)
Date of Event 06/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluation: product testing could not be performed since the product was not returned for evaluation as it was discarded.Therefore, the reported issue could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaint was received from this production order.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the an explant was planned.Additional information received revealed that explant was performed from patient's left eye and the original preloaded intraocular lens (iol) was replaced with another lens of different model but of same diopter.The patient was dissatisfied with the original lens that he had blurred visual acuity, halos and the letters were overlapping.The pre-operative visual acuity was 20/100 and post-implantation with the original lens, the distance vision improved to 20/20, however, the patient was unhappy with the near vision.After the explant and replacement with the new lens, the visual acuity was 20/20 -1.There was no other patient injury and the patient was very happy with the vision at day 1 post-operative visit.The original lens was discarded.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15676844
Report Number3012236936-2022-02616
Device Sequence Number12349515
Product Code MFK
UDI-Device Identifier05050474750609
UDI-Public(01)05050474750609(17)240713
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/13/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU165
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2022
Initial Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/13/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight100 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-