ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-13-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MX-13-T
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported unknown side "tissue expander deflated." device has been explanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Trend analysis: "review of all complaints of (b)(4): fluid leak for the period of oct 2020 through sep 2022 related to date opened indicates that 1 point is above the upper control limit.However, an additional analysis is not required because for (february/2021) no complaints were found in the last 24 months.No patterns were found; therefore, no additional actions are deemed required.The trend of (b)(4): fluid leak complaints will continue to be monitored and corrective actions will be taken in the future if deemed appropriate." further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Healthcare professional reported unknown side "tissue expander deflated." device has been explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: h3, h6 device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: observed an opening assessed as an unidentified (tear) opening (shell thickness whiting to spec) and a broken assessed as surgical damage.Additional observations: ¿ missing suture tan six on clock observed assessed as inconclusive.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a2, b5, d6a.
 
Event or Problem Description
Healthcare professional reported left side "tissue expander deflated." device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MX-13-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key15676916
Report Number9617229-2022-18677
Device Sequence Number6462941
Product Code LCJ
UDI-Device Identifier10888628044197
UDI-Public10888628044197
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/20/2025
Device Model Number133S-MX-13-T
Device Catalogue Number133S-MX-13-T
Device Lot Number3476947
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer11/02/2022
12/07/2022
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date11/24/2022
12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-