|
It was reported that, during a shoulder arthroscopy, the flush suture cutter had the tip at the bottom cracked and it was not cutting the suture thread.Surgery was resumed, without any surgical delay, with the same device.No further complications were reported.
|
|
H10: h3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A medical review states the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The provided pilgrim intake form was reviewed; however, it did not aid in determining the clinical root cause of the reported failure.The impact to the patient beyond that which has already been reported cannot be determined.Since there was no harm alleged to this patient and there were no other complications reported, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would re-assessed.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|