| Model Number |
IPN028446 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that air leaks on connection tubing.There was no reported injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#:(b)(4).The device history review could not be conducted since the lot number was not provided.Teleflex will continue to monitor and trend related events.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The device history record of batch number 74m2101787 has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.No nonconformance reports were originated for the lot in question that can be associated to the complaint reported.Dhr shows that the product was assembled & inspected according to our specifications.Customer complaint cannot be confirmed based only on the information provided, to perform an investigation and determine the source of defect reported it is necessary to evaluate the sample involved on this complaint.An attempt to duplicate the failure mode was made but at the time there is no inventory of the involved product code available at the facility nor is being manufactured at the time.If the sample becomes available this investigation will be updated with the evaluation results.
|
| |
|
Event or Problem Description
|
|
It was reported that air leaks on connection tubing.There was no reported injury.
|
| |
|