• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; TIBIAL BASEPLATE, SIZE 3

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; TIBIAL BASEPLATE, SIZE 3 Back to Search Results
Model Number TB-2203E-01M
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Scar Tissue (2060)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is in process.The device will not be returned for evaluation.When the investigation is complete, a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this event: #3013450937-2022-00305, #3013450937-2022-00306, #3013450937-2022-00307, #3013450937-2022-00308, #3013450937-2022-00309, #3013450937-2022-00310, #3013450937-2022-00311, #3013450937-2022-00312, #3013450937-2022-00313.
 
Event or Problem Description
It was reported that the patient was experiencing loss of mobility in their knee.The patient underwent a revision surgery on (b)(6) 2022.During the revision surgery, the surgeon identified scar tissue formation.The surgeon removed the scar tissue.The surgeon took out all of the implants besides the segmental stem, collar, locking ring, and tibial baseplate, and flashed these devices then re-implanted the devices.The surgeon scheduled another revision surgery on (b)(6) 2022 in order to revise some implants and perform a washout in order to prevent an infection.On (b)(6) 2022, the surgeon revised the following implants: distal femur axial pin, poly spacer, tibial hinge component, distal femur, and male-female midsection.No additional information regarding this event has been provided.
 
Event or Problem Description
It was reported that the patient was experiencing loss of mobility in their knee.The patient underwent a revision surgery on 06 october 2022.During the revision surgery, the surgeon identified scar tissue formation.The surgeon removed the scar tissue.The surgeon took out all of the implants besides the segmental stem, collar, locking ring, and tibial baseplate, and flashed these devices then re-implanted the devices.The surgeon scheduled another revision surgery on (b)(6) 2022 in order to revise some implants and perform a washout in order to prevent an infection.On (b)(6) 2022, the surgeon revised the following implants: distal femur axial pin, poly spacer, tibial hinge component, distal femur, and male-female midsection.No additional information regarding this event has been provided.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the adverse event could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
TIBIAL BASEPLATE, SIZE 3
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
PHILLIPS PRECISION INC
7 paul kohner pl
elmwood park NJ 07407
Manufacturer Contact
sara dailey
77 east halsey road
parsippany, NJ 07054
MDR Report Key15677481
Report Number3013450937-2022-00314
Device Sequence Number15033035
Product Code KRO
UDI-Device IdentifierB278TB2203E01M0
UDI-PublicB278TB2203E01M0
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K203815
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTB-2203E-01M
Device Catalogue NumberTB-2203E-01M
Device Lot Number89965-013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date10/26/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P/N 25000007E, ELEOS DISTAL FEMUR; P/N 25001140E, ELEOS MALE-FEMALE MIDSECTION; P/N 25001210E, ELEOS TIBIAL POLY SPACER; P/N 25002111E, ELEOS DISTAL FEMUR AXIAL PIN; P/N HC-15120-03M, ELEOS SEGMENTAL STEM; P/N HR-30000-03M, ELEOS COLLAR LOCKING RING; P/N KPONTP32E, ELEOS RESURFACING PATELLA; P/N PB-3200R-03M, ELEOS BIOGRIP HA COLLAR; P/N THSMWRS01M, ELEOS TIBIAL HINGE COMPONENT
Outcome Attributed to Adverse Event Hospitalization;
Patient Age70 YR
Patient SexFemale
-
-