BAXTER HEALTHCARE CORPORATION REVACLEAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114746L
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/28/2022
Event Type  malfunction  
Event Description
It was reported that during use with a revaclear 400 , an external blood leak at the top and bottom of the o-ring was observed.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
The device was received for evaluation.It was determined through testing that there was no external leakage of blood found on each dialyzer.Each dialyzer was found to have a yellow discoloration at the end of the fiber bundle.This yellow discoloration is not a internal or external leakage of blood inside the dialysate path, instead being an incorrect distribution of pur (polyurethane) being coated on the outside of the fibers.The reported condition of external leak was not verified.The cause is manufacturing related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - OPELIKA
1101 jeter ave
opelika AL 36801
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15677849
MDR Text Key303232532
Report Number3006552611-2022-00019
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414124076
UDI-Public(01)07332414124076
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K130039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number114746L
Device Lot NumberC422202301
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/03/2022
Initial Date FDA Received10/26/2022
Supplement Dates Manufacturer Received11/25/2022
Supplement Dates FDA Received11/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-