(b)(4).Batch # unk.Additional information was requested and the following was obtained: were the clips viewed endoscopically during the initial surgical procedure? yes how many days postoperative did the leak occur? i don't know for sure but patients usually be discharged the day following the surgery and was not because it has already presented nanormality in the immediate post operatory where was the leak anatomically located? majority doctors report cystic duct what was observed at the site of the leak upon reoperation? some doctors say that they don't even operate because they know that in the reoperation it is difficult to visualize exactly where it leakes so they put the drain and let it drain until it closes.Prior to loading, can you please confirm each of clip cartridges were placed in the lc800 base metal? i don't have information please confirm the brand and product code of the clip applier used with the lt300 clips? i have no information because i was not in surgery.Fact was potentially reported by the physician were malformed or scissored clips observed? if yes, please describe shape.Doctor does not report but they realize that, compared with another company, the lt 300 does not close well in the proximal position.When there is a malformation of the clip they clip again.Prior to use, did the or staff confirm alignment of the jaws of the clip appliers as requested in the ifu? yes.They always search to clip with aligned legs but there have been reports of cross clipping and having to clip again two other events were also reported, pc-001211673 and pc-001211675, were these all 3 separate patients and events? yea.These are other events and according to the doctors there are much more cases a fact that is making us molibize to change clips and follow procedures to recover customers is the surgeon experienced with the use of j&j clip appliers? yea.All doctors who reported have much experience in surgery and clip use did the surgeon experience any difficulties using the clip applier? if so, please describe.What is current patient status? some say that the jnj clipador is difficult to rotate, it is not ergonomic for rotation inside the cavity because it can't be faxed with one hand.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: it was stated that, " two other events were also reported, pc-001211673 and pc-001211675, were these all 3 separate patients and events? yea.These are other events and according to the doctors there are much more cases a fact that is making us molibize to change clips and follow procedures to recover customers" please confirm the specific number of patient events.Have any of these events been previously reported to ethicon? if yes, please provide complaint file number.For each patient event please provide: event date, product code and lot number involved, procedure, and event description.Was there any adverse patient outcome? if yes, was there any medical or surgical intervention performed? investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|