| Catalog Number |
03P79-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Hyperglycemia (1905); Elevated ketones/Diabetic Ketoacidosis (2364); Respiratory Acidosis (2482)
|
| Date of Event |
10/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022, abbott point of care was contacted by a customer regarding i-stat ec8+ cartridges that yielded suspected discrepant pco2 results on a male patient with severe acidosis.There was no patient information available at the time of this report.Date: (b)(6) 2022.Time: 13:59 14:27.Na: 126 mmol/l 125mmol/l.K: 8.4 8.2 mmol/l.Cl: 100 mmol/l 104 mmol/l.Tco2 6 <5.Urea: 19.6 mmol/l 19.5 mmol/l.Hct: 42%pcv 38% pcv.Ph: 6.772 6.833.Pco2: 4.55 kpa 2.33 kpa.Hco3: 5.0 mmol/l 2.9 mmol/l.Beecf:-30 <-30.Angap:30 mmol/l <>.Hb: 143 g/l 129 g/l.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on 23-nov-2022.A review of the device history record (dhr) confirmed the cartridge lot met finished goods release criteria.Retained testing met the acceptance criteria found in q04.01.003 rev.Ak, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been determined for ec8+ cartridge lot k22085.
|
| |
|