| Model Number |
CREP G2 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received a questionable creatinine plus ver.2 result for 1 patient sample on a cobas 8000 c701 module.The initial result was reported outside of the laboratory.The physician questioned the result.A new sample was collected from the patient.The reporter is looking for a medical explanation for such a decrease in the results.It was alleged that the difference in results between the two samples was not possible due to the patient¿s condition.On (b)(6) 2022 , the initial result of the initial sample was 134 umol/l.On (b)(6) 2022 , the result of the new sample was 77 umol/l.The c701 module serial number was (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The calibration data provided from 03-sep-2022 to 13-sep-2022 was within specifications.The alarm trace showed multiple sample short alarms.The preanalytical checklist showed that the centrifugation condition was 2200 g for 11 minutes.According to the manufacturer's recommendation, the centrifugation conditions should be 1300-2000 g for 10 minutes.The investigation determined the event was due to a preanalytic issue at the customer site.Preanalytics are within the customer's responsibility.There was no indication of a device malfunction.
|
| |
|