MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 40MM M SPEC+12; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US 12/14 ARTICUL 40MM M SPEC+12; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-08-000
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Hemorrhage/Blood Loss/Bleeding (1888); Unspecified Infection (1930)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Ppf and implant records received.Ppf alleges bone fracture and infection.It is unknown which bone was fractured.Doi: (b)(6) 2010 - dor: (b)(6) 2012 (left hip).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2012 the patient had left mom hip revised, a +12 delta ceramic femoral head was placed during this procedure along with competitor cup, screws, and polyliner.It may be assumed that the delta ceramic femoral head was depuys.On (b)(6) 2014 the patient had left hip revision with cage/cup construct to address the preoperative diagnosis of aseptic loosening of left acetabular component (competitor component).The indications for surgery included: increasing pain and ¿screws were broken¿.The patient has also had dislocations, which were noted to be ¿consistent with migration of the cup¿.During the surgery, the surgeon observed competitor screw was broken.The femoral head was revised along with the cup/liner/screws.A +12 revision ceramic head was placed.Medical records from (b)(6) 2022 note patient has left hip pain.Patient has been receiving injections from a pain doctor on a monthly basis (hip/back pain).Unknown etiology.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned to depuy synthes for evaluation.All available x-rays were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found.The reported allegation cannot be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12/14 ARTICUL 40MM M SPEC+12
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15680224
Report Number1818910-2022-21394
Device Sequence Number12372921
Product Code LPH
UDI-Device Identifier10603295033028
UDI-Public10603295033028
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2012
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1365-08-000
Device Catalogue Number136508000
Device Lot Number2946196
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer10/10/2022
11/07/2022
02/23/2023
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date11/02/2022
11/08/2022
02/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
12/14 ARTICUL 40MM M SPEC+12; CORAIL AMT COLLAR SIZE 12; PINNACLE MTL INS NEUT40IDX56OD; PINNACLE SECTOR II CUP 56MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexMale
-
-