MEGADYNE MEDICAL PRODUCTS, INC. MEGASOFT UNIVERSAL PLUS Back to Search Results
Catalog Number 0847
Medical Device Problem Codes Arcing of Electrodes (2289); Unintended Electrical Shock (4018)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 10/27/2022 only event year known: 2022.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No serial number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: were there any patient consequence related to this event? if yes please describe.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported during a spine procedure the bovie pencil was arcing.The staff in the room placed a sticky pad on the patient and unplugged the megasoft pad.They noticed when the anesthesiologist touched the patient's hand, it delivered a shock.The patient was not in contact with metal as far as they know.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 11/29/2022.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 12/6/2022 investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the 0847 pad was returned with no apparent damage.The pad was connected to the generator and no anomalies were noted.The event described could not be confirmed as the device was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of the ethicon endo-surgery quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch gs21004207, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEGASOFT UNIVERSAL PLUS
Common Device Name
MEGASOFT UNIVERSAL PLUS
Manufacturer (Section D)
MEGADYNE MEDICAL PRODUCTS, INC.
4545 creek road
cincinnati OH 45242
Manufacturer (Section G)
MEGADYNE MEDICAL PRODUCTS, INC.
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key15680505
Report Number1721194-2022-00095
Device Sequence Number20448863
Product Code GEI
UDI-Device Identifier10614559104842
UDI-Public10614559104842
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K133726
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date10/31/2023
Device Catalogue Number0847
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer11/18/2022
12/04/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date11/29/2022
12/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1