| Model Number |
5DCD |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Additional information was requested and the following was obtained: no patient consequence, another device has been used for the procedure.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, the jaws were very flexible.The clamp was found to be defective and could not complete the intervention with it.No patient consequence.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 12/06/2022.D4: batch # x94t20.Investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the 5dcd instrument was returned with no damage in the external components.The instrument was tested and functioned properly as it did open and closed without any difficulties noted.The end effectors were inspected and no broken parts were noted.Although no conclusion could be reached on the cause of the reported event.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.The event described could not be confirmed as the instrument performed without any difficulties noted.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot x94t20 number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|