It was reported that during the surgery, after fired, noted the staples malformed.Changed another two to continue the surgery, the same problem happened again.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
|
(b)(4).Batch #284a49.Investigation summary : the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pce60a device was returned with no apparent damage and with three ecr60b reloads(b,c and d) present.The reload(b) was received fully fired, and the reload(c) was received fully fired with damage on reload deck and drivers.The reload(d) was received unfired.In addition, the knife was noted to be nicked, a b-formed staple was found in knife channel.The device was tested for functionality in the straight position with reload(d) and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.One possible cause for this type of damage to the knife is when the device is fired over an already existing staple line, hard object or thicker tissue than indicated.Repeatedly firing across existing staple lines can also reduce the ability to cut cleanly.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.Please reference the instruction for use for more information.The damage to the reload is consistent with the device being clamped over a hard object.To mitigate the potential for staples getting into the reload and interfering with the knife path during device firing, prior to reloading the device, rinse the anvil and reload jaw in sterile solution and then wipe the anvil and reload jaw to clean any formed but unused staples from the device.Additionally, proper care should be taken when placing the device on the tissue to be stapled, to ensure that no hard obstruction such as a clip is included with the tissue inside the jaws.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot and batch number, and no non-conformances were identified.
|