|
|
| Model Number |
NONB8STF |
| Medical Device Problem Codes |
Leak/Splash (1354); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: nonb8stf nonb8stf no blade 8mm std fix, lot #: j2d2556y.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic nissen fundoplication, during the entire procedure, the trocar did not hold pnuemo when a 5mm instrument was inserted.The device leaked air and can be heard the entire case.The surgeon had to switch out trocars because the device could not keep the abdomen inflated.Pulled out the 5mm grasper to try and let the abdomen fill up with pneumo again so that the surgeon could continue with the procedure.When the trocar seal leaked, the abdomen deflated and the surgeon cannot performed the procedure, it came to a halt and the surgeon had to wait again to inflate the abdomen before proceeding.The surgeon continued to open a new 8mm trocars to complete the procedure.Some device leaked more than the others.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|
|
|