• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number N/A
Medical Device Problem Codes Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the nurse found the curve of the needle has a leak.It was immediately changed.No patient injury reported.It was reported this occurred with two devices.This report addresses the second device.
 
Additional Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of refn3628 showed two other similar product complaint(s) from this lot number.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample analysis, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of leaking infusion sets was confirmed and the cause appeared to be manufacturing-related.The product returned for evaluation was two 22ga x 0.75¿ bard winged infusion sets with y-sites.Usage residues were observed throughout both samples.An attempt to infuse water through the sample using a 12ml syringe revealed both samples to be patent; however, both samples exhibited leaks at the needle exit sites.During manual manipulation, both needles swiveled easily within their housings.The needles were easily removed from the tubing/housing, requiring little effort.Microscopic inspection of the needle shafts revealed adhesive residue surrounding the needle shafts.Some gaps were observed in the adhesive coverage.The leakage at the needle exit site appeared to be caused by poor bonds between the needle shafts and the tubing/housings.The adhesive residue suggested that adhesive was applied during device assembly; however, both the gaps in the adhesive and the weakness of the bond between the needles and other device components suggested that a strong bond was not formed during device assembly.Photographs have been forwarded to the manufacturing site for further evaluation.
 
Event or Problem Description
It was reported the nurse found the curve of the needle has a leak.It was immediately changed.No patient injury reported.It was reported this occurred with two devices.This report addresses the second device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGED INFUSION SET WITH INJECTION SITE 22G X 0.75IN
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15681837
Report Number3006260740-2022-04156
Device Sequence Number6336813
Product Code FMI
UDI-Device Identifier00801741065873
UDI-Public(01)00801741065873
Combination Product (Y/N)N
Initial Reporter CountrySA
PMA/510(K) Number
K863606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Model NumberN/A
Device Catalogue Number2205222
Device Lot NumberREFN3628
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/09/2022
Supplement Date Received by Manufacturer01/23/2023
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date01/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-